Probe Setup and Use¶
This section explains probe hardware, how to wire probes to your current source, and how to position and contact a probe against your device under test (DUT). It applies to probes used with EZ500 and NT200 Series Microsanj systems.
Probe Parts¶
Each probe assembly consists of the following parts:
- X, Y, Z control knobs – fine-positioning controls used to move the probe tip in three axes
- Probe arm – the body of the probe assembly connecting the control knobs to the probe tip
- Probe tip – the fine point that makes electrical contact with the DUT
- Magnetic plate – the base of the probe assembly, which magnetically holds the probe to the (ferrous) probe station surface
- Probe BNC – the BNC connector at the end of the probe cable used to connect the probe to the current source
[diagram/image in source — not reproduced here: labeled photo of a probe assembly showing the X/Y/Z control knobs, probe arm, probe tip, magnetic plate, and probe BNC connector]
BNC Connectors¶
The Probe BNC connector carries two conductor paths:
- Center conductor — goes to the probe tip
- Shield/ground — goes to shield/GND
To wire a probe to a current source, you will typically need one or more BNC-to-banana adapters, since current sources commonly terminate in banana jacks rather than BNC connectors. The source material shows the following adapter types used in various combinations:
- Female BNC to Male Banana
- Male BNC to T-Splitter
- Male BNC to Female Banana
- Female BNC to Single Male Banana
On the probe side, the BNC connector's center conductor (probe tip / positive) and shield (GND / negative) are wired using these adapters to the current source's positive and negative terminals.
[diagram/image in source — not reproduced here: photos of each BNC/banana adapter type with pin callouts]
Connecting Probes to Current Source¶
Connect the probe to your current source using BNC-to-banana adapters:
- Connect the current source's negative terminal to the shield/GND path of the probe BNC connector, using a Female BNC-to-Male Banana adapter (or equivalent).
- Connect the current source's positive terminal to the probe tip path of the probe BNC connector, using the appropriate BNC/banana adapter and, where needed, a Male BNC T-splitter to combine connections.
- Confirm polarity at the probe BNC connector before applying current: the center conductor path corresponds to the probe tip (positive), and the shield path corresponds to GND (negative).
[diagram/image in source — not reproduced here: two wiring diagrams showing negative/positive terminal connections through banana-to-BNC adapters and a T-splitter to the probe BNC connector]
Probe Positioning¶
Position the probe on the probe station before bringing it near the DUT:
- Center the travel range of the X, Y, and Z controls, then place the probe onto the probe station.
- Tilt the probe to the side so that only the edge of the magnetic plate is touching the probe station.
- Slide the probe along the probe station and position it so the probe tip is close to the area on the DUT you wish to test.
[diagram/image in source — not reproduced here: three-step illustration of probe placement and sliding on the probe station]
Contacting the Probe to the DUT¶
Once the probe tip is positioned near the DUT:
- Touch the probe tip to the area of interest using the X, Y, and Z controls.
CAUTION: Go slow and watch the probe tip carefully. Once contact is made, damage can occur quickly.
[diagram/image in source — not reproduced here: diagram of probe tip contacting the DUT with polarity (-/+) marked]
EZ500 Probe Station Probe Connections¶
The source material includes an additional reference diagram showing probe connections on an EZ500 probe station, with two probes contacting a DUT (marked with – and + polarity).
[diagram/image in source — not reproduced here: EZ500 probe station photo with probe connections and DUT polarity labeled]